5. Select TAB as the delimiter and click Finish

NOTE: If you want to save graphed data points to Microsoft Excel, you can use the computer mouse to right click on the

graphed data and select “Export to Microsoft Excel”.

Clicking the DOWNLOAD button next to the “DOWNLOAD FACTORY DEFAULTS” label downloads the original default

controller settings, including ramp/soak program values. You can review these settings by clicking on the INDEX scroll

arrows which will then show the DESCRIPTION of the setting and its VALUE. These default settings cannot be changed.

CONTROL MODE BOX

This section allows adjustment of the primary set modes, normal set modes, proportional, integral, derivative values, and

manual output level.

The controller SET MODE can be set to one of three control modes:

1. NORMAL SET—the controller will maintain either a single set point temperature or a fixed percentage output

power, or it can send an analog signal for a programmable power supply determined by the NORMAL SET MODES

selection:

i.

PID—this allows the controller to automatically control to a single set point temperature. The controller

can be tuned for control stability by adjusting the “PROPORTIONAL BANDWIDTH”, “INTEGRAL GAIN”, and

“DERIVATIVE GAIN” values in the PID PARAMETERS section. See 3.0 Controller Tuning above for

information on how to tune the controller. See also the options in the OUTPUT BOX for additional details.

ii.

MANUAL—this allows the controller to provide a fixed percentage of output power. The output power can

be set by adjusting the scale pointer in the MANUAL OUTPUT LEVEL section or by entering a percentage

output level in the box. Generally, a negative value will cause the TE device to cool, and a positive value will

cause the TE device to heat.

iii.

ANALOG OUT—this allows the controller to be used with an external, variable-voltage programmable DC

power supply to provide true analog control (not pulse width modulated control). The main difference in

ANALOG OUT mode is the output of the H-bridge is forced to always be either +100% or -100% depending

if the controller is calling for heating or cooling. There is no pulse-width modulation of the voltage that is

put into JP7-3 and JP7-4. The H-bridge circuitry within the controller is used strictly to control the polarity

of the voltage to the thermoelectric device, and is not used to vary the magnitude of that voltage. The