CTS-M17 Installation, Operation and Maintenance Manual

B84450-001-000 RG

May 12, 2005

5

The keypad is a set of three push buttons recessed along the upper right side of the

enclosure. It is not necessary to open the cover for use of the keypad. Access to the

menus is password restricted. Press any key for 3 seconds to enter the menus (see

below for details).

1.3 Password

The factory preset default password is 0017. The password can be changed.

IMPORTANT: RECORD PASSWORD IN A SECURE PLACE. If the

password is lost, the unit must be returned to

QEL to be reset.

1.4 Output Signals

Options are 4-20 mA or 2-10 VDC linear. Choose the option by moving the jumper

on JP1. On over range concentrations the signal can exceed the range slightly.

IMPORTANT: Voltage monitoring should be into a monitor with

at least 50,000 Ohms input impedance. Voltage

monitoring is not recommended over long

distances, as these signals are more susceptible

to induced noise than current signals

1.5 Meter Jacks

Test pads, accepting standard test probes are provided on the circuit card. These

are labeled SIGA, SIGB and COMMON.

To test the analog output signal, measure:

SIGA to COMMON shorts to internal ground.

SIGA to SIGB without interrupting signal to monitor.

To Test voltage signals measure:

SIGB to COMMON.

Figure 2 - Output Signal Select Jumper (JP1)