Pacific Research Solutions RI-1 User Manual
Page 50

Pacific Research Solutions
RI-1 and PE-1 User Manual
Page 44
S-Cmd 58 REMOTE SOFTWARE RESET
This command allows a method for resetting the controller remotely, just the same as if you were to do a power on reset.
SYNTAX:
58
STORAGE:
N/A
PARAMETERS: Hard coded 19
DEFAULT:
N/A
EXAMPLE: 58 19 Resets the controller software only.
S-Cmd 59 REMOTE EEPROM INITIALIZE
This command allows you to remotely reset your EEPROM (Non-Volatile Memory) data with factory defaults.
SYNTAX:
59
STORAGE:
N/A
PARAMETERS: Hard coded 86.
DEFAULT:
86
EXAMPLE: 59 86 Initialize the EEPROM only.
S-Cmd 60 RESERVED
S-Cmd 61 RESERVED