beautypg.com

ONICON F-4000 Series User Manual

Page 34

background image

11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax +1 (727) 442-5699 • [email protected]

F-4000 Series Ultrasonic Flow Meter Manual 05/15 - 0707-13 / 18838

Page 34

14.

15.

Display will integrate zero flow for 60 seconds, do
not hit enter until complete.

16.

None

Zero flow integration complete.

17.

You are no longer in the program mode.
The meter is now measuring flow.

4.4 NAVIGATING THE RUN MODE MENU PAGES

The display contains two lines of alphanumeric characters. Each line can be independently set to
display any of the seven operating parameters listed below.

Flow Rate – displays the current instantaneous flow value (e.g. 20.3 GAL/M). Note that reverse
flow is indicated by a minus sign preceding the displayed value (e.g. – 20.3 GAL/M).

Flow Total – displays the current flow total (e.g. 15030.5 GAL) Note: flow multipliers can be kilo
(K) or mega (M). The resolution of the displayed total can be set to show 1 – 3 digits to the right of
the decimal point. Totalizer will roll over to zero when the maximum count is exceeded.

Gain Level Percentage (Valc) – displays the signal amplifier gain level as a percentage value. The
higher the Valc value, the better. Valc values greater than 30 are generally considered good.

Aeration Percentage (Vaer) - displays the noise level generally associated with entrained air as a
percentage value. Lower values indicate less entrained air.

Signal Strength (mV) - displays the actual signal level in millivolts. Higher signal levels indicate a
good sonic coupling.

Keypad Pushbuttons

Description of functions in run mode

Not used in run mode.

Right arrow activates cursor below left most character on the
display. The cursor must be active and visible before you can
change the displayed data. Left arrow not used.

Used to scroll up and down through the display menu options.