CET™ Scale, MTE Top Mount Configuration

Installation By Mounted Applicator Tool

The correct end of the tape to insert is shown
below.

5.

Applicator Tool / CET Scale Orientation

Tape Scale orientation arrows

Tool orientation
arrows

Both Tape Scale arrows, and Applicator Tool
orientation arrows need to point in the same
direction, while the tape is being applied.

In the example above, the side of the tape scale
that needs to go into the tool entry channel first
is called the “correct” end.

Tape entry
channel

“Correct” end of tape to
insert into applicator tool.

While mounted in the MTE sensor’s fixture, CET
Scale is fed through the entry channel and pressed
onto the mounting surface with the spring loaded
contact cylinder.

The contact cylinder has two positions,

Up and

Down, controlled by the black position knob on the
side of the Applicator Tool.

Up, allows clearance for feeding of the CET Scale
through the entry slot into the “start” position.
Down, locks the tape scale onto the “start” position.
Moving the Applicator Tool along in the

Down

position will automatically press the CET Scale to
the mounting surface.

4.

Applicator Tool - bottom view

CET Scale
adhesive surface

CET Scale clear adhesive
backing, unravelling

Spring loaded
contact cylinder

6.

Install the Applicator Tool into the MTE Mount
Tighten the mounting screws.