beautypg.com

Acquisition setup, Acquisition setup……7 – Measurement Computing DaqView for DaqTemp Applications User Manual

Page 9

background image

DaqTemp Applications

967996

1108-0902, rev 2.0

DaqView, pg. 7

Acquisition Setup

Selecting the second tab of the main window displays the Acquisition Setup window. The four main parts of
this window include setup parameters for triggering and configuring the scan. These settings will be used
when an acquisition-to-disk is started by selecting “Acquire” (last item on the toolbar). When the trigger is
satisfied, the scans are collected at the selected scan frequency and stored to disk in the designated file.

Acquisition Tab Selected

Acquisition Setup Window

Parameter

Description

Pre-Trigger

The number of scans to acquire before the trigger event.

Trigger Event

Selects the trigger source, depending on the device. Possible trigger sources include:

Immediately

– arms and executes the trigger immediately.

Key Hit

– arms the acquisition and waits for the user to press a key.

External TTL Rising/Falling

– waits for a rising or falling edge slope on P1-25.

Hardware Rising/Falling Edge

– monitors value with hysteresis on selected channel; triggers

when parameter is satisfied.

External TTL High/Low

– waits for a TTL high or low level on pin 25 of connector P1.

Hardware Above/Below Level

– monitors value on selected channel; triggers when parameter is

satisfied.

Above/Below Level

– monitors value on selected channel; triggers when parameter is satisfied.

Rising/Falling Edge

– monitors value with hysteresis on selected channel; triggers when

parameter is satisfied.

Inside/Outside Window

– monitors upper and lower values on selected channel; triggers when

parameters are satisfied.

Digital Pattern

– monitors 8-bit pattern on selected digital input channel; triggers when

parameters (less/greater than or equal to/not equal to) are satisfied.