beautypg.com

Specifications, Analog input, Chapter 4 – Measurement Computing USB-7202 User Manual

Page 19

background image

19

Chapter 4

Specifications

All specifications are subject to change without notice.
Typical for 25 °C unless otherwise specified.
Specifications in italic text
are guaranteed by design.

Analog input

Table 1. Analog input specifications

Parameter

Conditions

Specification

A/D converter type

16-bit successive approximation type

Number of channels

8 single-ended

Input configuration

Individual A/D per channel

Sampling method

Simultaneous

Absolute maximum input voltage

CHx IN to GND

±15 V max

Input impedance

100 M

Ω min

Input ranges

Software-selectable

±10 V, ±5 V, ±2 V, ±1 V

Sampling rate

Scan to system memory

0.6 S/s to 50 kS/s, software-selectable

BURSTIO scan to 32 k sample FIFO

20 S/s to 50 kS/s, software-selectable

Throughput

Software paced

500 S/s all channels, system-dependent

Scan to system memory (Note 1)

(100 kS/s) / (# of channels); maximum of
50 kS/s for any channel

BURSTIO scan to 32 k Sample FIFO

(200 kS/s) / (# of channels), maximum of
50 kS/s for any channel

Resolution

16 bits

No missing codes

15 bits

Crosstalk

Signal DC–25 kHz

–80 dB

Trigger source

Software-selectable

External digital: TRIG_IN

Calibration

Cal factors stored in firmware. Cal factors
must be applied to data via application
software.

Note 1:

Maximum throughput scanning to system memory is machine dependent.

Table 2. Calibrated absolute accuracy

Range

Accuracy (mV)

±10 V

5.66

±5 V

2.98

±2 V

1.31

±1 V

0.68

Table 3. Accuracy components - All values are (±)

Range

% of Reading

Gain error at FS (mV)

Offset (mV)

±10 V

0.04

4.00

1.66

±5 V

0.04

2.00

0.98

±2 V

0.04

0.80

0.51

±1 V

0.04

0.40

0.28