beautypg.com

Channel-gain queue – Measurement Computing E-1608 User Manual

Page 14

background image

E-1608 User's Guide

Functional Details

14

Leave unused input channels either floating or tied to an AGND terminal. Source impedances should be kept as
small as possible to avoid settling time and accuracy errors.

Figure 4 shows DIFF channels 0-3 connected to a ground path resistor.

Figure 4. DIFF connections with ground path resistor

Channel-Gain queue

The channel-gain queue feature allows you to configure a list of channels, modes, and gains for each scan. The
settings are stored in a channel-gain queue list that is written to local memory on the device.

The channel-gain queue list contains one or more channel numbers, modes, and range settings. You can
configure up to 8 elements. The channels can be listed in any order, and can include duplicate channels for
sampling at different ranges.

An example of a 4-element list is shown in the table below.

Sample channel gain queue list (SE mode)

Element

Channel

Range

0

CH5

BIP5V

1

CH1

BIP10V

2

CH3

BIP1V

3

CH5

BIP5V

Carefully match the gain to the expected voltage range on the associated channel or an over range condition
may occur. Although this condition does not damage the device, it does produce a useless full-scale reading,
and can introduce a long recovery time due to saturation of the input channel.

For more information about analog signal connections

For more information about analog input connections, refer to the Guide to DAQ Signal Connections (this
document is available on our website at

www.mccdaq.com/signals/signals.pdf

).