Signal connections, Analog input, Floating voltage source – Measurement Computing E-1608 User Manual

Page 13

background image

E-1608 User's Guide

Functional Details

13

The E-1608 pinout locations are shown in Figure 3 below.

Figure 3. E-1608 pinout

Signal connections

Analog input

You can configure the analog inputs for SE or DIFF mode. The input voltage range is software selectable for
±10 V, ±5 V, ±2 V, or ±1 V.

With SE mode, connect up to eight inputs to

CH0x

to

CH3x

. SE mode requires two wires:

 Connect one wire to the signal you want to measure (

CH#x

).

 Connect one wire to the analog ground reference (

AGND

).

With DIFF mode, connect up to four differential inputs to

CH0H/CH0L

to

CH3H/CH3L

. DIFF mode requires

two wires plus a ground reference:

 Connect one wire to the high/positive signal (

CHxH

).

 Connect one wire to the low/negative signal (

CHxL

).

 Connect one wire to the analog ground reference (

AGND

).

Floating voltage source

When connecting DIFF voltage inputs to a floating voltage source, make sure the DIFF input channel has a DC
return path to ground. To create this path, connect a resistor from each low channel input to an AGND pin. A
value of approximately 100 kΩ can be used for most applications.