CMG-5TD Digital Accelerometer

Quick start

The screw on/off cover for output offset adjustment on the CMG-5TD
top cap.

The serial number on the label, on the top or side of the instrument.

3.2

Initial check-out

This section gives a quick outline for initial system test, which should
performed prior to installation. You must provide 10-36 V DC power and a

PC running Scream.

Attach the levelling baseplate to the base of the CMG-5TD using the screws

provided. Set the CMG-5TD onto a flat surface. Using the large hex screws on
the baseplate in conjunction with the bubble level, level the instrument.

Connect the CMG-5TD directly to the PC using the supplied power/data cable

Switch on the power supply. Using the PC, start Scream and take the

following steps:

1. Click on 'Setup', from the ‘File’ menu

2. Select the ‘Com Ports’ tab

3. Click on the COM port to which the instrument is connected.

4. The factory-set baud rate for the CMG-5TD is 19200. Select 19200 from

the list.

5. Click on the ‘OK’ button to return to the main Available Streams

window.

MAN-050-0005

7

Issue D - April 2013