beautypg.com

2 frequency response curves – Guralp Systems CMG-5TD User Manual

Page 30

background image

CMG-5TD Digital Accelerometer

Calibration

G is the acceleration output sensitivity (gain constant) of the
instrument. This relates the actual output to the desired input over the

flat portion of the frequency response.

A is a constant which is evaluated so that A × H (s) is dimensionless

and has a value of 1 over the flat portion of the frequency response. In
practice, it is possible to design a system transfer function with a very

wide-range flat frequency response.

The normalising constant A is calculated at a normalising frequency

value fm = 1 Hz, with s = j fm, where j = √–1.

H (s) is the transfer function of the sensor, which can be expressed in

factored form:

In this equation, z

n

are the roots of the numerator polynomial, giving

the zeros of the transfer function, and p

m

are the roots of the

denominator polynomial giving the poles of the transfer function.

In the calibration pack, G is the sensitivity given on the first page, whilst the

roots z

n

and p

m

, together with the normalising factor A, are given in the Poles

and Zeros table. The poles and zeros given are measured directly at Güralp

Systems' factory using a spectrum analyser.

6.1.2 Frequency response curves

The frequency response of the CG-5TD is described in the normalised
amplitude and phase plots provided. The response is measured at low and

high frequencies in two separate experiments. Each plot marks the
low-frequency and high-frequency cut-off values (also known as –3 dB or

half-power points).

MAN-050-0005

30

Issue D - April 2013