beautypg.com

3 calibration with scream – Guralp Systems CMG-5TD User Manual

Page 32

background image

CMG-5TD Digital Accelerometer

Calibration

calculated using a Fourier transform. Because the input signal has
predominantly low-frequency components, this method generally gives

poor results. However, it is simple enough to be performed daily.

Injecting a sinusoidal current of known amplitude and frequency

allows the system response to be determined at a spot frequency.
However, before the calibration measurement can be made, the system

must be allowed to reach a steady state; for low frequencies, this may
take a long time. In addition, several measurements must be made to

determine the response over the full frequency spectrum.

Injecting white noise into the calibration coil gives the response of the

whole system, which can be measured using a spectrum analyser.

You can perform calibration using the built-in CMG-DM24 digitizer, which

can generate step and sinusoidal calibration signals.

6.3

Calibration with Scream!

Calibration is most easily done using a PC running Güralp's Scream! Software.
In this section, broadband noise calibration will be used to determine the

complete sensor response in one action. Please refer to the CMG-DM24 and
Scream manuals for information on other calibration methods.

1. In Scream!'s main window, right-click on the digitiser's icon and select

Control.... Open the Calibration pane.

2. Select the calibration channel corresponding to the instrument, and

choose Broadband Noise. Select a suitable duration and amplitude,
and click Inject now. A new data stream, ending Cn (n = 0 – 7) or MB,
should appear in Scream!'s main window containing the returned

calibration signal.

MAN-050-0005

32

Issue D - April 2013