CMG-5TD Digital Accelerometer

Calibration

If you want to repeat the calibration to obtain more precise values at a

particular frequency of interest, or to check that a sensor is still functioning
correctly, you can inject calibration signals into the system using a Güralp

digitizer or your own signal generator, and record the instrument's response.

6.1.3 Obtaining copies of the calibration pack

Our servers keep copies of all calibration data that we send out. In the event
that the calibration information becomes separated from the instrument, you

can obtain all the information using our free e-mail service. Simply e-mail

[email protected]

with the serial number of the instrument in the subject

line, e.g.

From: [email protected]

To: [email protected]
Subject: T5215

The server will reply with the calibration documentation in Microsoft Word
format. The body of your e-mail will be ignored.

6.2

Calibration methods

Three common calibration techniques are available:

Injecting a step current allows the system response to be determined in
the time domain. The amplitude and phase response can then be

MAN-050-0005