Ecg performance testing – Fluke Biomedical PS420 User Manual
Page 40
PS420
Users Manual
30
Code Display
Selects:
117
CAL WAVES Calibrated cardiac output
waves.
118 NONCAL
WVE
Uncalibrated cardiac output
waves. Gives 4 different
values per 3, 5, and 7 L/min
waves.
119
CAL 1 SEC
Calibration pulse of 1 degree
for 1 second.
120
CAL 4 SEC
Calibration pulse of a delta of
402 Ohms for 4 seconds.
121
SET BT 37C 37 °C (98.6 °F)
122
SET BT 36C 36 °C (95.9 °F)
When the monitor under test indicates that it is ready to
perform a Cardiac Output calculation, press ENTER on
the Simulator to generate a test. The 4-second calibration
pulse provides a delta change of 402 Ohms. This
provides a standard for testing cardiac output units.
If you select any cardiac output waveform key a second
time while a waveform is proceeding, the selected
waveform will stop. The blood temperature will return to
37 °C (98.6 °F.)
ECG Performance Testing
Square Wave
Code
Display Selects:
128
2 Hz SQR
2.0 Hz square waveform.
129
.125 Hz SQ
0.125 Hz square waveform.
Triangle Wave
Code
Display Selects:
130
2 Hz TRIAN
2.0 Hz triangle waveform.
131
2.5 Hz TRI
2.5 Hz triangle waveform
Pulse Wave
Code
Display Selects:
132
PULSE 30
Pulse of 30 BPM, width of 60 ms.
133
PULSE 60
Pulse of 60 BPM, width of 60 ms.