Pacemaker pulse amplitudes, lead ii, Pacemaker pulse width – Fluke Biomedical PS420 User Manual
Page 37

Multiparameter Simulator
Simulating Functions
27
Pacemaker Pulse Amplitudes, Lead II
When you select any pacemaker waveform, you can
adjust and set pulse amplitudes in Lead II. These can be
set through direct code entry, as below, or selected and
adjusted through first selecting 64 = PACE AMP. After
selecting 64, press ENTER, and then use the CHANGE
keys
to scroll to the available presets.
Code
Display
Sets pacemaker pulse
amplitude (Lead II) of:
224
PACE 2 mV
2 mV
225
PACE 4 mV
4 mV
226
PACE 6 mV
6 mV
227
PACE 8 mV
8 mV
228
PACE 10 mV
10 mV
Pacemaker Pulse Width
When you select any pacemaker waveform, you can
adjust and set pulse width. These can be set through
direct code entry, as below, or selected and adjusted
through first selecting 65 = PACE WIDTH. After selecting
65 press ENTER, and then use the CHANGE keys
to scroll to the available presets.
Code
Display
Sets pacemaker pulse
width of:
229
PACE 0.1mS
0.1 ms
230
PACE 0.5mS
0.5 ms
231
PACE 1 mS
1.0 ms
232
PACE 1.5mS
1.5 ms
233
PACE 2 mS
2.0 ms