Accessories – Fluke Biomedical PS420 User Manual
Page 14
PS420
Users Manual
4
Accessories
Item Part
Number
Standard Accessories
Users Manual CD-ROM
2631721
Users Manual (printed)
2631808
9 VDC Battery Eliminator
2647372
Optional Accessories
Cardiac Output Adapter Box
2462200
Temperature Cable *
Blood Pressure Cable *
Cardiac Output Cable *
* Contact your local Fluke Biomedical Sales Agent for further details
ECG
12 Lead ECG with nine independent outputs referenced to RL
Baseline Impedances..........................................500, 1000, 1500 or 2000 Ohms for Leads I, II, and III
High Level Output ...............................................1000 x (Lead II)
Rates...................................................................30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, and 300 BPM
Default Rate ........................................................80 BPM.
Rate Accuracy.....................................................± 1% of selection
Adult or Pediatric Waveform
ECG Amplitudes..................................................0.5, 1.0, 1.5, and 2.0 mV
Amplitude Accuracy ............................................± 2 %. (Lead II)
Superimposed Artifact.........................................50 and 60 Hz, muscle, baseline wander, and respiration