Ecg/arrhythmia, Adult and pediatric nsr qrs – Fluke Biomedical PS420 User Manual
Page 31
Multiparameter Simulator
Simulating Functions
21
ECG/Arrhythmia
The Simulator replicates 37 different types of arrhythmias,
from inconsequential types of PNCs to asystole. In
addition, the Simulator can send waveforms to test any
electrocardiograph, and can accommodate twelve-lead
configurations with independent outputs for each signal
lead referenced to the right leg (RL).
Adult and Pediatric NSR QRS
An adult NSR with a QRS width of 80 ms, or a pediatric
NSR with a QRS width of 40 ms, can be set. These will
remain in effect throughout ECG and arrhythmia
selections until changed by reentering the below codes
and pressing ENTER.
Code Display
Selects:
11
PEDIATRIC
Pediatric NSR with QRS width of
40 ms.
12
ADULT
Adult NSR with QRS width of 80
ms.
NSR
The Simulator replicates fifteen normal sinus rhythms, or
NSRs. These NSRs can be set through direct code entry,
as below, or selected and adjusted through first selecting
9 = NSR. After selecting 9, press ENTER, and then use
the CHANGE keys
to scroll to the available presets.
Code
Display
Selects NSR rate of:
170
NSR 30BPM
30 BPM
171
NSR 40BPM
40 BPM
172
NSR 60BPM
60 BPM
173
NSR 80BPM
80 BPM
174
NSR 100BPM 100 BPM
175
NSR 120BPM 120 BPM
176
NSR 140BPM 140 BPM
177
NSR 160BPM 160 BPM
178
NSR 180BPM 180 BPM
179
NSR 200BPM 200 BPM
180
NSR 220BPM 220 BPM
181
NSR 240BPM 240 BPM
182
NSR 260BPM 260 BPM
183
NSR 280BPM 280 BPM
184
NSR 300BPM 300 BPM