Cardiac output option, Respiration, Temperature – Fluke Biomedical PS420 User Manual
Page 18

PS420
Users Manual
8
Cardiac Output Option
Catheter Type ........................................................Baxter-Edwards, 10 cc
Blood Temperature ...............................................37
°C (98.6 °F) and 36 °C (95.9 °F)
CO for 2 Degrees C ...............................................3, 5, and 7 L/Min
CO for 20 Degrees C .............................................3, 5, and 7 L/Min
Cal Pulse ................................................................Of 1 degree C for 1 second; of delta 402 Ohms for 4 seconds
Accuracy ................................................................± 5 %
Computational Constant
2 Degrees C ........................................................0.561
20 Degrees C ......................................................0.608
Left to Right Shunt * .............................................2 and 20 Degrees C
Faulty Injectate *....................................................2 and 20 Degrees C
*Note: These four CO simulations are examples of defective (uncalibrated) curves.
Respiration
Baseline Impedances............................................500, 1000, 1500, 2000 Ohms (Leads I - III)
Lead Selection.......................................................LL or LA
Impedance Variations ...........................................0.2, 0.5, 1.0, and 3.0 Ohms
Impedance Accuracy ............................................± 5 %
Rates ......................................................................0 (apnea), 15, 20, 30, 40, 60, 80, 100, 120 BPM
Apnea .....................................................................12 sec, 22 sec, 32 sec, and continuous
Rate Accuracy .......................................................± 2 %
Temperature
30
°C (86 °F), 35 °C (95 °F), 37 °C (98.6 °F), 40 °C (104 °F), 42 °C (107.6 °F), compatible with YSI 400/700 series
Accuracy ................................................................± 0.25 Degree C