Fluke Biomedical PS420 User Manual
Page 39

Multiparameter Simulator
Simulating Functions
29
Code Display
Selects:
107
3L/M @ 2C
Cardiac output wave of 3
L/min for 2 degrees.
Computational constant is
0.561.
108
5L/M @ 2C
Cardiac output wave of 5
L/min for 2 degrees.
Computational constant is
0.561.
109
7L/M @ 2C
Cardiac output wave of 7
L/min for 2 degrees.
Computational constant is
0.561.
110
3L/M @ 20C Cardiac output wave of 3
L/min for 20 degrees.
Computational constant is
0.608.
111
5L/M @ 20C Cardiac output wave of 5
L/min for 20 degrees.
Computational constant is
0.608.
Code
Display Selects:
112
7L/M @ 20C
Cardiac output wave of 3
L/min for 20 degrees.
Computational constant is
0.608.
113
FLT INJ 2
Faulty injection cardiac output
wave for 2 degrees.
Computational constant is
0.561.
114
L- T SHT 2
Left to right shunt cardiac
output wave for 2 degrees.
Computational constant is
0.561.
115
FLT INJ 20
Faulty injection cardiac output
wave for 20 degrees.
Computational constant is
0.608.
116
L- R SHT 20
Left to right shunt cardiac
output wave for 20 degrees.
Computational constant is
0.608.