Blood pressure – Fluke Biomedical PS420 User Manual
Page 17

Multiparameter Simulator
Specifications
7
Blood Pressure
Input/Output Impedance .......................................350 Ohms
Exciter Input Limit .................................................± 10 V
Exciter Input Frequency Range ...........................DC to 4000 Hz
Transducer Sensitivity..........................................5 or 40 µV/V/mmHg
Level Accuracy ......................................................± 1% ± 1 mmHg
Static Levels
BP 1.....................................................................-10, 0, 80, 160, 240, 320, and 400 mmHg
BP 2.....................................................................-10, 0, 50, 100, 150, 200, and 240 mmHg
Channel Selections
Arterial 120/80 .....................................................Channels 1 and 2
Radial Artery 120/80............................................Channels 1 and 2
Left Ventricle 120/00 ...........................................Channels 1 and 2
Right Ventricle 25/00 ...........................................Channels 1 and 2
Central Venous 15/10..........................................Channel 2
Pulmonary Artery 25/10.......................................Channel 2
Pulmonary Wedge 10/2.......................................Channel 2
Left Atrium 14/4 ...................................................Channel 2
Automatic Swan/Ganz......................................... Every 20 seconds
Manual Swan/Ganz ............................................ Changes every time entry is selected
Synchronized with all normal sinus rates
Physiologically tracks all arrhythmia selections