Impedance Variations
The Simulator replicates impedance variations of 0.2, 0.5,
1.0, 2.0, and 3.0 and 2000 delta Ohms. Use the LA/LL
switch on the top panel to select which lead will have the
respiration signal on it. This will not affect the baseline
operation.
The variation can be set through direct code entry, as
below, by keying the preset code and pressing ENTER.
You can also use the SCROLL keys
to cycle
through the other presets before pressing ENTER.
Code
Display
Selects respiration delta Ohms:
161
R DELTA .2
0.2 Ohms
162
R DELTA .5
0.5 Ohms
163
R DELTA 1
1.0 Ohms
164
R DELTA 2
2.0 Ohms
165
R DELTA 3
3.0 Ohms
Apnea
The Simulator replicates the general apnea types by
duration only: continuous; and 12, 22, and 32 seconds.
Apnea can be set through direct code entry, as below, by
keying the preset code and pressing ENTER. You can
use the SCROLL keys
to cycle through the other
presets before pressing ENTER.
Code
Display
Selects respiration apnea:
144
APNEA ON
ON. Respiration rate of 0 RPM.
145
APNEA OFF
OFF. Respiration reverts to
previously selected respiration
rate.
146
12 SEC APN
For 12 seconds.
147
22 SEC APN
For 22 seconds.
148
32 SEC APN
For 32 seconds.