Fluke Biomedical IDA-4 Plus User Manual
Operators manual, Ida-4 plus
                        Table of contents                                                                        
                Document Outline
- IDA-4 Plus Infusion Pump Analyzer Operators Manual
 - 1. Introduction
 - 2. An Overview
 - 3. Operating Instructions 
- Connecting Infusion Pumps under Test 
- General Operating Notes
 - First Time Use 
- Report Header
 - Set Clock
 - Test Parameters
 - Printer
 - Measuring Circuit
 - Test Fluid
 - Warning of Contamination
 - Start-up Screen
 - Status All Channels Screen
 - Utilities
 - Recall Tests
 - Set Clock
 - LCD Set-up
 - Printer Options
 - Report Header
 - RS232 Port
 - Engineering Tests
 - Test Parameters
 - Channel Set-up Menu
 - Device Information
 - Sub Menu1
 - Sub Menu2
 - Flow Test Screen (Prime Mode)
 - Flow Test Screen (Start Mode)
 - Flow Test Screen (Measuring Mode)
 - Flow Test Screen (End of Test Mode)
 - Flow Graph Screen
 - Occlusion Test Screen
 - Occlusion Test Screen (Wait Mode)
 - Occlusion Test Screen (Start Mode)
 - Occlusion Test Screen (Measuring Mode
 
 
 
 - Connecting Infusion Pumps under Test 
 - 4. Maintenance, Service and Calibration
 - Appendices
- A. Infusion Devices Delivery Methods
 - B. Computer Control Commands 
- Overview
 - Controlling the IDA-4 with HydroGraph™
 - Controlling the IDA-4 Plus Using Serial Protocol
 - Command Conventions
 - Perform a Flow Test [CnF,a,b,c]
 - Perform a Volume Test [CnV,a,b,c]
 - Perform an Occlusion Pressure Test [CnO, a,b,c]
 - Perform a PCA Test [CnPCA,a,b,c]
 - Perform a Dual Flow Rate Test [CnD,a,b,c,d,e,f]
 - Fetch the Current Flow Measured on a Channel [FLOWn]
 - Fetch the Current Volume Accumulated on a Channel [VOLn]
 - Fetch the Current Pressure on a Channel [PRESn]
 - Fetch the Peak Pressure on a Channel [PKPRESn]
 
 
 
