Operating instructions, Connecting infusion pumps under test, Chapter 3 – Fluke Biomedical IDA-4 Plus User Manual
Page 23

3-1
Chapter 3
Operating Instructions
This chapter discusses and illustrates the method
of connecting units under test and the
various operational modes.
Connecting Infusion Pumps under Test
See Inlet Hose Connections and Outlet Hose Connections drawings.
The Inlet Hose Connection drawing on the next page shows a method of connecting the 
inlets from units under test together with a priming syringe, both connected via a 3-way 
tap. 
This method is the only method that should be employed with this instrument.
The Outlet Hose Connection drawing on page 3 shows the drain hose connections from the 
unit to an open drainage collection vessel. Please note that the drain Fluid Outlets should 
each have an independent hose to the drainage vessel and should not be joined in any way. 
Note: Infusion device measurement accuracy is also affected by the 
compliance of the infusion device and delivery tubes. This is because 
there is a momentary blockage of flow when the measurement 
transducer is emptied. The back-pressure created by these events 
could cause some of the flow to be diverted into the compliance of the 
infusion device. For most infusion devices, the effect is trivial (less 
than 0.5%), but with very soft connecting tubing and large air pockets 
in the device or tubing, it could be significant. The tubing between the 
infusion device and the IDA-4 Plus should therefore be fairly rigid 
and all air should be removed before starting each flow test. At high 
flow rates it may be desirable to increase the compliance of the inlet 
tubing (longer or softer tube) if the back pressure interferes with the 
occlusion sensing alarms in the infusion device.
