Report header, Rs232 port – Fluke Biomedical IDA-4 Plus User Manual
Page 31

Infusion Pump Analyzer
Operating Instructions
3
3-9
Report Header
The REPORT HEADER of three lines and up to 28
characters on each line is entered via this screen. The
entry is stored and will appear on all printouts, it should
also be noted that this header can not be changed once a
test is in progress. Entry is made via the alphanumeric
matrix using the 5 button keypad, a bar code reader or if
connected, via an external keyboard. On subsequent
selection of this screen the existing stored entry is
shown. This is stored in non-volatile memory and
retained when the instrument is switched off.
To use the alphanumeric matrix, use the direction keys
to move the cursor to the desired character, and select
by pressing the ENTER key.
Continue until header is complete.
Moving the cursor to any OK position accepts the text,
and move to the next field or screen. Selecting DEL
deletes a character, and the ESC character escapes back
to the previous field or screen
RS232 Port
To set the RS232 protocol, use the left and right keys to
select the item to be changed then the up and down keys
to change to the required setting.
The options are:
Speed: 150 - 38400 Baud
Parity: ODD - EVEN – NONE
Data: 7 or 8 bits
Stop: 1or 2 bits
When correct settings are achieved press ENTER to
accept these settings which are stored until next
changed via this screen.
U t i l i t y R e p o r t H e a d e r
………………………………………………………………..
………………………………………………………………..
………………………………………………………………..
O K 0 1 2 3 4 5 6 7 8 9 D E L
O K A B C D E F G H I J
O K K L M N O P Q R S T E S C
O K U V W X Y Z . , / -
O K
R S 2 3 2 P O R T
S p e e d P a r i t y D a t a S t o p
1 9 2 0 0 N O N E 8 1
U s e t h e l e f t a n d r i g h t a r r o w
K e y s t o s e l e c t .
U s e t h e u p a n d d o w n a r r o w
Keys to change.
P r e s s E N T E R t o A c c e p t