Fluke Biomedical IDA-4 Plus User Manual
Page 43

Infusion Pump Analyzer
Operating Instructions
3
3-21
Occlusion Test Screen (End of Test Mode)
This view of the OCCLUSION TEST screen allows
the operator to save, print or delete the results of the
occlusion test.
Save
This saves the results of the occlusion test, or the series
of tests, to non-volatile memory for later printing or
transfer to a computer via the RS232 port.
Save & Print
This saves the results as described above, then sends a
report to a printer connected to the printer port on the
instrument. The report will be in either full or summary
format, as previously selected by the operator from the
UTILITIES / PRINTER menu.
This sends a report to a printer connected to the printer
port on the instrument. The report will be in either full
or summary format, as previously selected by the
operator from the UTILITIES / PRINTER menu.
Status
This switches to the STATUS ALL CHANNELS
screen to allow the operator to use another channel or
change settings via the UTILITIES menu.
Graph
This switches to the OCCLUSION GRAPH screen for
the current channel.
Delete
This deletes the current set of results from memory
without saving to non-volatile memory. CAUTION:
Once this option has been invoked there will not be any
way to recover, view or print the results of this test.
C H A N N E L 1 O C C L U S I O N
P e a k P r e s s u r e
0 . 0 0 p s i
Time of Peak
00: 00: 00
Total Test Time
00: 00: 00
Set Flow Rate
ml /h
SAVE
SAVE & PRINT
GRAPH
STATUS
DELETE