Perform a volume test [cnv,a,b,c, Perform an occlusion pressure test [cno, a,b,c, Perform a pca test [cnpca,a,b,c – Fluke Biomedical IDA-4 Plus User Manual
Page 71

Infusion Pump Analyzer
Computer Control Commands
B
B-3
Perform a Volume Test [CnV,a,b,c]
This command starts a flow / volume test on the specified channel. The IDA-4 Plus status
all channels screen will prompt the operator to prime the channel. Once primed, the
operator will have to press the Start key on the IDA-4 Plus front panel or send the [CnV]
command again.
Parameters:
n = channel number
a = control number of device under test
b = technician name
c = set flow rate
Perform an Occlusion Pressure Test [CnO, a,b,c]
This command starts an occlusion pressure test on the specified channel. The IDA-4 Plus
status all channels screen will prompt the operator to start the test. The operator can either
press the start key on the front panel or send the [CnO] command again.
Parameters:
n = channel number
a = control number of device under test
b = technician name
c = set flow rate
Perform a PCA Test [CnPCA,a,b,c]
This command starts the PCA test on the specified channel. The IDA-4Plus status all
channels screen will prompt the operator to prime the channel and start the test. Once
primed, the operator can press the start key on the front panel or send the [CnPCA]
command again.
Parameters:
n = channel number
a = control number of device under test
b = technician name
c = set basal flow rate