Occlusion test screen, Occlusion test screen (wait mode), Pressure – Fluke Biomedical IDA-4 Plus User Manual
Page 41

Infusion Pump Analyzer
Operating Instructions
3
3-19
Occlusion Test Screen
This is the screen that allows the operator to control the
occlusion test. It has a menu across the bottom of the
screen that changes to reflect the progress of the test. It
is not necessary to prime the transducer to perform an
occlusion test.
The operator sets the units of pressure displayed on this
screen from the TEST PARAMETERS SCREEN.
Occlusion Test Screen (Wait Mode)
When the OCCLUSION TEST screen is first entered it
puts the transducer into a Tare measuring mode (i.e.
subtracts the ambient pressure). While the transducer is
in this mode the WAIT prompt is displayed.
Wait
This menu option is never active, it informs the operator
to wait until the transducer has taken a tare reading.
When the tare reading has completed this, the menu
option changes to START, and the instrument advances
to the start mode. The pump should be stopped or
running very slowly during the wait period.
Status
This invokes the STATUS ALL CHANNELS screen.
ESC
Selecting ESC or pressing the ESC key on the external
keyboard takes the operator back to the CHANNEL
SETUP menu.
C H A N N E L 1 O C C L U S I O N
0.00
PRESSURE
psi
TIME
00 : 00 : 00
Peak Pressure
psi
Time of Peak 00: 00: 00
WAIT
STATUS ESC