Flow 1, Volume 0.00, Dual rate test screen – Fluke Biomedical IDA-4 Plus User Manual
Page 53: Dual rate test screen (prime mode), Prime, Status, Dual rate test screen (start mode), Auto-start, Start

Infusion Pump Analyzer
Operating Instructions
3
3-31
Dual Rate Test Screen
This is the screen that allows the operator to control the
dual rate flow test. It has a menu across the bottom of
the screen that changes to reflect the progress of the
test.
Dual Rate Test Screen (Prime Mode)
When the screen is first displayed the operator is
prompted to prime the channel. This is essential to
ensure that the measuring circuit is filled with fluid.
Prime
This prompt stays on the screen while there is any air in
the measuring circuit. Pressing ENTER while this
prompt displays PRIME will not have any effect. As
the circuit is primed the prompt changes to START.
There may be some flickering between PRIME and
START as bubbles percolate through the system.
Status
This invokes the STATUS ALL CHANNELS screen.
ESC
Selecting ESC or pressing ESC on the external
keyboard takes the operator back to the CHANNEL
SETUP menu.
Dual Rate Test Screen (Start Mode)
When the measuring circuit is primed successfully the
screen advances to the Start Mode.
Auto-Start
Pressing ENTER on this option puts the IDA-4 Plus
into AutoSTART mode. The prompt will flash until the
instrument detects that the fluid is flowing, then start
the test.
Start
When the channel is primed this option is displayed and
enabled. Pressing ENTER while this item is selected
starts the test and advances to the Measuring Mode.
Status
This invokes the STATUS ALL CHANNELS screen.
ESC
Selecting ESC or pressing ESC on the external
keyboard takes the operator back to the CHANNEL
SETUP menu.
C H A N N E L 1 D U A L F L O W
FLOW 1
0.00
ml /h
VOLUME 0.00
ml
FLOW 2
0. 00
VOLUME
E l a p s e d T i m e
00: 00: 00
I n s t . F l o w
0. 00 ml /h
B a c k P r e s s u r e
0. 00 psi
PRIME
STATUS ESC
C H A N N E L 1 D U A L F L O W
FLOW 1
0.00
ml /h
VOLUME 0.00
ml
FLOW 2
0. 00
VOLUME
E l a p s e d T i m e
00: 00: 00
I n s t . F l o w
0. 00 ml /h
B a c k P r e s s u r e
0. 00 psi
AutoSTART
START STATUS
ESC