Channel set-up menu – Fluke Biomedical IDA-4 Plus User Manual
Page 33

Infusion Pump Analyzer
Operating Instructions
3
3-11
Channel Set-up Menu
This screen allows the operator to select the type of test
to be performed as a rapid test or elect to enter further
details before selecting the type of test. The layout of
options on this menu has been optimized to reduce the
number of key presses when using the rapid test
options.
Using the arrow keys, move the cursors to the required
selection and press ENTER to advance to the selected
sub menu.
The following menu options are available:
Occlusion
This advances to the OCCLUSION TEST screen
where the operator will be prompted through the
occlusion test. It will not be possible to save the test
results.
Flow
This advances to the FLOW TEST screen where the
operator will be guided through Flow tests. It will not
be possible to save the test results.
Dual Flow
This advances to the DUAL FLOW INFORMATION
screen, even for a rapid test as some information is
required to perform this test. It will not be possible to
save the test results.
Dev Info
This advances to the DEVICE INFORMATION
screen where information about the device being tested
will be requested. If test results are to be saved, this
option must be selected.
PCA
This advances to the PCA TEST INFORMATION
screen to request information required to perform a
PCA test. It will not be possible to save the test results.
ESC
This returns the operator to the STATUS ALL
CHANNELS screen. The ESC key on an external
keyboard will have the same effect.
C H A N N E L 1 S E T U P
OCCLUSION DEV
INFO
FLOW
PCA
DUAL FLOW
ESC