Ecg-performance-testing, Respiration, Invasive blood pressure – Fluke Biomedical ProSim 4 User Manual
Page 39: Non-invasive blood pressure

Vital Signs Simulator
Detailed Specifications
29
ECG-Performance-Testing
Amplitude .............................................................. 1 mV. Other leads are proportional to Lead II (reference lead) in
percentage per:
Lead I .............................................................. 70
Lead II ............................................................. 100
Lead III ............................................................ 30
Lead V1 .......................................................... 24
Lead V2 .......................................................... 48
Lead V3 .......................................................... 100
Lead V4 .......................................................... 120
Lead V5 .......................................................... 112
Lead V6 .......................................................... 80
Square Wave ......................................................... 60 ms at 2.0 Hz
Respiration
Rate ........................................................................ 0 (OFF), 10 to 100 BrPM in 10 BrPM steps
Impedance Variations (
Δ
Ω
) ................................. 1
Ω
Accuracy Delta ......................................................
±
(10 % + 0.05 ohm)
Baseline ................................................................. 500
Ω
to circuit common, giving 1000
Ω
between any two leads
Accuracy Baseline ................................................
±
5 %
Respiration Lead ................................................... LA or LL (default)
Invasive Blood Pressure
Channels ................................................................ 1 electrically isolated from all other signals
BP Output .............................................................. Circular DIN 5-Pin
Input/output Impedance ....................................... 300
Ω
±
10 %
Exciter Input Range .............................................. 2.0 to 16.0 V peak
Exciter-Input Frequency Range ........................... DC to 5000 Hz
Transducer Sensitivity ......................................... 5
μ
V/V/mmHg
Pressure Accuracy ...............................................
±
(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation
only
Static Pressure ...................................................... 0, 80, 160, and 250 mmHg
Dynamic Waveforms
Synchronization .................................................. To ECG heartrate
Chambers simulated systolic/diastolic pressure:
Type
IBP (Arterial)
IBP (Left Ventrical)
Adult 60/30
60/0
Adult 120/80
120/0
Adult 150/100
150/0
Adult 200/150
200/0
Neonatal 35/15
35/0
Neonatal 70/40
70/0
Non-Invasive Blood Pressure
Pressure Units ....................................................... mmHg
Manometer (Pressure Meter)
Range ................................................................. 10 to 400 mmHg
Resolution ........................................................... 0.1 mmHg (for display purposes)
Accuracy .............................................................
±
(1 % reading +1 mmHg)
Pressure Source ................................................... Inflation bulb or device under test