Invasive blood pressure simulation – Fluke Biomedical ProSim 4 User Manual
Page 28

ProSim™ 4
Users Manual
18
Note
When the battery discharges to a low level threshold, all NIBP
functions may be disabled.
Invasive Blood Pressure Simulation
The Product can simulate an invasive blood pressure transducer. To do an
invasive blood pressure simulation, connect the Product to the monitor as shown
in Figure 13.
ProSim 4
Patient
Monitor
IBP Cable
gne021.eps
Figure 13. Invasive Blood Pressure Connections
To access the invasive blood pressure parameters, touch the IBP control in the
Home screen to show the screen in Figure 14 in the display.
gne006.bmp
Figure 14. Invasive Blood Pressure Screen
The invasive blood pressure, heart rate, respiration, and the non-invasive blood
pressure parameters are shown at the top of the display. To change the invasive
blood pressure, touch one of the pre-defined blood pressure controls in the
display.
To zero the simulated output, touch the Zero softkey.
To set a static IBP pressure signal to the IBP monitor, touch the Static Pressure
softkey from the IBP screen. The static pressure screen in Figure 15 shows in
the display.