Arrhythmia simulation – Fluke Biomedical ProSim 4 User Manual
Page 20

ProSim™ 4
Users Manual
10
Heart rate, respiration, invasive blood pressure, and non-invasive blood pressure
parameters are shown at the top of the screen. To change the heart rate, touch
one of the pre-defined heart rate controls in the display.
The ECG function can be set to an adult or neonatal ECG signal. To change
between Adult and Neonatal, touch the softkey control in the lower-left corner of
the display. The softkey text will change to the one value not set. For example,
when Neonatal is shown in the softkey, an adult ECG signal is simulated.
Touch the Home softkey to go back to the Home screen.
Table 8 shows the percentage of the signal amplitude value that is put on each
ECG lead.
Table 8. ECG Lead Amplitudes
Waveform I
II
III
V1
V2 V3 V4 V5 V6
Normal Sinus & Performance
70 % 100 %
30 % 24 % 48 % 100 %
120 % 112 %
80 %
Arrhythmia Simulation
The Product can simulate a number of arrhythmias on the ECG leads. As well as
physiological waveforms, the Product can supply a signal to measure the
performance of an ECG monitor. To do an arrhythmia simulation or do a monitor
test with a performance wave, touch the Arrhythmia control in the Home screen.
Page 1 of the screens in Figure 4 shows in the display.
Page 1
Page 2
gne008.bmp
Figure 4. Arrhythmia Screens