Fluke Biomedical ProSim 4 User Manual
Page 31

Vital Signs Simulator
Autosequences
21
Push to abort the step and move to the subsequent step. When the
autosequence has moved to step two, shows in the display. Touch to
pause the step. shows in the display when the auto sequence is paused.
Push to continue the step for the time left when the step was paused.
To abort the autosequence, touch the Stop softkey and go back to the
autosequence view screen.
Table 9 is a list of autosequences that are built into the Product.
Table 9. Autosequences
Autosequence
Sequence Steps
Run Time
Cardiac Failure
ECG 80 bpm, IBP Off, NIBP Off, and Respiration Off
00:45
ECG PVC (LV), IBP Off, NIBP Off, and Respiration Off
00:30
ECG VTach (LV) 160 bpm, IBP Off, NIBP Off, and Respiration Off
00:30
ECG VFib Coarse, IBP Off, NIBP Off, and Respiration Off
00:30
ECG Asystole, IBP Off, NIBP Off, and Respiration Off
00:15
STOP Total
Time
02:30
xercise
ECG 60 bpm, IBP Off, NIBP Off, and Respiration Off
00:30
ECG 90 bpm, IBP Off, NIBP Off, and Respiration Off
00:30
ECG 120 bpm, IBP Off, NIBP Off, and Respiration Off
00:30
ECG 150 bpm, IBP Off, NIBP Off, and Respiration Off
00:30
ECG 90 bpm, IBP Off, NIBP Off, and Respiration Off
00:30
ECG 60 bpm, IBP Off, NIBP Off, and Respiration Off
00:30
REPEAT Total
Time
03:00
Respiration
ECG 60 bpm, IBP Off, NIBP Off, and Respiration 110 brpm
00:30
ECG 60 bpm, IBP Off, NIBP Off, and Respiration 60 brpm
00:30
ECG 60 bpm, IBP Off, NIBP Off, and Respiration 20 brpm
00:30
ECG 60 bpm, IBP Off, NIBP Off, and Respiration 0 brpm (Apnea) 00:12
REPEAT Total
Time
01:42
Monitor Testing
ECG 120 bpm, IBP 200/150 (Art), NIBP 200/150 (167), and
Respiration 40 brpm
01:30
ECG 60 bpm, IBP 120/80 (Art), NIBP 120/80 (93), and Respiration
20 brpm
01:30
ECG 40 bpm, IBP 60/0 (LV), NIBP 60/30 (40), and Respiration
10 brpm
01:30
STOP Total
Time