Autosequences – Fluke Biomedical ProSim 4 User Manual
Page 29

Vital Signs Simulator
Autosequences
19
gne007.bmp
Figure 15. Static Pressure Screen
To set the simulated pressure, touch one of the pre-defined static pressure
controls.
Touch the Home softkey to go back to the Home screen.
Autosequences
Autosequences are a series of steps that change the output of the Product
automatically. For example, to do a monitor test, you could set normal ECG then
wait some time. Then touch Hypertensive and wait more time, then touch
Hypotensive and wait. The monitor testing autosequence does these changes
for you automatically. Each step of the monitor testing autosequence sets
simulation parameters and after the allotted time period, it does the subsequent
step.
To do an autosequence:
1. Touch the More softkey if it is shown in the Home screen.
2. Touch one of the autosequence controls to show the autosequence steps in
the display. The Monitor Testing autosequence is shown in Figure 16.
gne012.bmp
Figure 16. Monitor Testing Autosequence Screen