Ecg simulation – Fluke Biomedical ProSim 4 User Manual
Page 18

ProSim™ 4
Users Manual
8
Table 7. Pre-Defined Patient Simulations
Simulation Name
Parameter Pre-Set
Value
Normal
Wave Form
NSR (Adult)
ECG Rate
60 bpm
Respiration Rate
20 bpm
IBP Channel 1
120/80 mmHg (Art)
NIBP
120/80 (93) mmHg
Hypertensive
Wave Form
NSR (Adult)
ECG Rate
120 bpm
Respiration Rate
40 bpm
IBP Channel 1
200/150 mmHg (Art)
NIBP 200/150
(167)
mmHg
Hypotensive
Wave Form
NSR (Adult)
ECG Rate
40 bpm
Respiration Rate
10 bpm
IBP Channel 1
60/0 mmHg (LV)
NIBP
60/30 (40) mmHg
Control of heart rate, respiration, invasive blood pressure, and non-invasive
blood pressure simulations are also accessed through the controls on page 1 of
the Home screen. A number of arrythmias are set through the Arrhythmia
control. Refer to the applicable section to learn more about each simulation
function.
To show page 2 of the Home screen, push the More softkey. Page 2 shows
controls that access autosequences, pressure and leak tests, and all setup
parameters. To go back to page 1 of the Home screen, push the Back softkey.
ECG Simulation
The Product simulates normal heart signals (ECG) as well as a variety of heart
arrhythmias. You set adult or neonatal simulation and change the heart rate
(beats per minute) through the ECG control.
To measure the ECG performance of a monitor, connect the Product to the
monitor as shown in Figure 2. A maximum of ten ECG leads can be connected to
the Product.