Fluke Biomedical ProSim 4 User Manual
Page 30

ProSim™ 4
Users Manual
20
The autosequence above, shows the sequence steps. Some autosequences
stop at the last step. The screen also shows that the length of time to complete
the autosequence is 3 minutes and the length of time for each step is
60 seconds.
Touch the Home softkey to go back to the second page of the Home screen.
3. Touch the Start softkey to start the sequence and show the screen in
Figure 17 in the display.
gne013.bmp
Figure 17. Monitor Testing Autosequence Steps Screen
The top part of the display shows the simulation parameters and their values.
These parameter values update at each autosequence step. The middle part of
the display shows which step the autosequence is on and how much time is left
for the step. The sequence timer shows the length of time to complete all steps of
the autosequence. If this is an autosequence that repeats, the sequence timer
resets to total sequence time when the sequence starts step one.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)