Fluke Biomedical RF303 User Manual
Page 59
Operation
Remote Operation via the RS-232 Function
3
3-23
Table 3-2. Available Duplex Mode Commands
Command Description
Parameters
Required
Return Value
(Response)
GOTOREMOTE
Places the Analyzer in
the duplex remote
mode.
None *
GOTOLOCAL
Returns the Analyzer to
normal (local mode)
operation / exits duplex
remote mode.
None
*
IDENT
Identifies the Analyzer
model and firmware
version number.
None
Model Analyzer name
and the version of the
firmware.
Example: MODEL
Analyzer VERSION
X.XX
Where X.XX is the
version number.
VER
Identifies the firmware
version number.
None
Firmware version
number.
Example: VERSION
X.XX
Where: X.XX is the
version number.
SETMODE:{parm}
Changes the format of
the Analyzer response
to the RDMETER
command. Separate the
command from the
desired mode
(parameter) using a
colon. Refer to
RDMETER below.
Example:
SETMODE:OUTPUT
OUTPUT (for
measuring
generator
output).
RFLKG (for
measuring
HF leakage).
RARF (for
conducting
CQM test).
*
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)