Fluke Biomedical RF303 User Manual
Page 41
Operation
Testing
the
ESU
3
3-5
2. Connect the ESU Neutral/Dispersive electrode to the left blue jack
dispersive, using the test lead, Part # 2772171.
3. Continue to load selection.
To connect the ESU to the Analyzer for a bipolar output test:
1. Connect the ESU Bipolar-1 electrode to the yellow jack (marked) active,
using one of the test leads from the set, Part # 1903307.
2. Connect the ESU Bipolar-2 electrode to the left blue jack dispersive, using
the other test lead from the set, Part # 1903307.
3. Continue to load selection.
To select the desired test load resistance:
1. Press the
OHMS SELECT
increase (+) or decrease (-) buttons until the
test load value you want appears in the display. After approximately one
second, the display returns to a blank screen. Below is an example of a test
load value as it appears in the display.
WATTS
mA
OHMS
ENT
MODE SELECT
OHMS SELECT
exz03.eps
Note
The default (or initial power-up) test load selection is 300 ohms.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)