Fluke Biomedical RF303 User Manual
Page 42
RF303
Operators Manual
3-6
To measure the ESU’s power output:
1. Activate the ESU.
2. If the
WATTS
indicator lamp on the panel is lit, go to Step 3. If it is not
lit, press the
MODE SELECT
(
ENT
) button to select
WATTS
and go to
Step 3.
3. View the power measurement on the display. Here is an example:
WATTS
mA
OHMS
ENT
MODE SELECT
OHMS SELECT
exz04.eps
To measure the ESU’s current output:
1. While the ESU is still activated, press the
MODE SELECT
(
ENT
) button
until the milliamperes (
mA
) indicator lamp lights.
2. View the ESU’s current output (in mA). Below is an example.
WATTS
mA
OHMS
ENT
MODE SELECT
OHMS SELECT
exz05.eps
To end the generator output test:
Deactivate the ESU.
W
Caution
To avoid damage to the Analyzer and the ESU under test,
do not exceed the Analyzer’s duty cycle, which is 30
seconds ON during any continuous one-minute period.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)