Fluke Biomedical RF303 User Manual
Page 58
RF303
Operators Manual
3-22
•
All responses are uppercase, followed by CR and LF.
•
Measurement readings from the Analyzer are returned identically as
they would appear on the Analyzer's display; that is, as numbers, with
or without a decimal point, and followed by any appropriate unit of
measurement.
•
Errors always return a response beginning with: ?ERR=XX (plus the
description). Table 3-1 lists each of the error responses.
Table 3-1. Error Responses
Error Response
Explanation
?ERR=01,BAD OR MISSING
PARAMETERS
The received command requires
parameters, and there were none, or
they were not recognized.
?ERR=02,ILLEGAL NUMBER OF
PARAMETERS
The command did not include the
correct number of parameters.
?ERR=03,PARAMETER OUT OF
RANGE OR SYNTAX ERROR
One or more of the parameters were
out of the correct range, or there was
an error in the syntax of the
command.
?ERR=04,COMMAND NOT
AVAILABLE FROM THIS MODE
The Model Analyzer is not in the
special sub mode within the remote
that is required for the command.
?ERR=05, DEACTIVATE ESU
BEFORE ADJUSTING
There is an ESU signal present at the
meter inputs that is not allowed for
the command to be executed.
?ERR=06,ACCESS DENIED
The command is a calibration
command and the calibration jumper
is not connected
Table 3-2 describes various commands available in Duplex Mode.