beautypg.com

Specifications, Maintenance, Calibration – Fieldpiece SMAN4 - 4 Port Wireless Digital Manifold User Manual

Page 2: Limited warranty, Obtaining service, Pulling a vacuum, Set vacuum alarms, Sman4 to hvac guide® (hg3), Pressure, Microns for vacuum

background image

11

16

12

17

13

18

14

19

15

20

Pulling a Vacuum

Follow all manufacturers’ evacuation procedures over those in

this manual.

1 Connect your SMAN4 in-line with your vacuum pump and the

system, then power on your SMAN4.

2 Draw a vacuum on the system. The SMAN4 will automatically

sense the negative pressure and display the vacuum in microns.

3 Press Alarm to activate the High (Hi) alarm or Low (Lo) alarm. No

display of Lo or Hi means alarm is off. Note: The stopwatch will

start when an alarm is activated. Pressing Alarm will reset the

stopwatch.

4 When the rate of change in pressure is less than 1 micron per 10

seconds, Stable will appear to the left of the micron reading.

Additional Tips:

To achieve a deep vacuum as quickly as possible:

1 Use the shortest hoses with the largest diameter available.

2 Remove schrader cores and core depressors. Core removal tools

like the "MegaFlow Valve Core Removal Tool" can be purchased

from Appion, Inc. to help with this process.

3 Inspect the rubber seals at both ends of your hoses for damage

that may result in leakage.

4 Do not use hoses with low loss fittings when evacuating or

pulling a vacuum on a system.

Set Vacuum Alarms

1 Hold ALARM for one second to enter Alarm Set mode.

2 Press ALARM to toggle between Alarm Hi and Lo.

3 Use UP or DOWN ARROW to select the value and press ENTER

to lock in each digit. Once last digit in Alarm Lo is locked, your

SMAN4 will automatically revert to normal testing mode.

SMAN4 to HVAC Guide® (HG3)

Your SMAN4 can wirelessly send measurements to the HG3 System

Analyzer (FW v 3.42 or higher) for deeper analysis and data logging. Your

SMAN4 can send suction and liquid line pressures and temperatures,

superheat/subcooling calculations, indoor wet bulb and outdoor drybulb

temperatures with target superheat calculation (if SMAN4 is wirelessly

receiving these measurements), and vacuum micron measurements

(for datalogging) all in real-time. From the HG3, you can print reports

on a PC to hand to customers and/or file for your own record-keeping.

The HG3 System Analyzer can receive wireless measurements from

the SMAN4 to the following tests: CheckMe!®, Superheat, Subcooling,

Charging Jacket, and Data Logger.

1 Connect your SMAN4 to the system for measurement of suction

and liquid line pressures and temperatures.

2 Sync wireless devices to SMAN4 to receive indoor wet bulb and

outdoor dry bulb measurements. See Target Superheat section for

wireless IDWB and ODDB instructions.

3 Press circular button on SMAN4 to enter Superheat/Subcooling, T1

T2 Direct, or Saturation temperature modes.

4 Hold the SYNC button until a beep is heard (>1 sec). Wireless

signal bars icon in lower right corner of display will appear and

begin searching.

5 Set HG3 to one of the following tests: CheckMe!®, Superheat,

Subcooling, Charging Jacket, or Data Logger.

6 With HG3, highlight a test line that corresponds to one of the

measurements being sent from the SMAN4.

7 Hold SYNC button on HG3 until search pattern shows on right of

test line. All receivable real-time measurements from the SMAN4

will populate corresponding test lines on HG3.

8 Press OUTPUT on HG3 to view deeper analysis of the system.

Note: The SMAN4 and HG3 must be within 10 feet (3m) for initial

sync to occur. Once synced, the wireless range is 100 feet (30m).

Specifications

Mini-USB port: For updating to newer versions of firmware

Display size: 5 inches (diagonal)

Backlight: Blue (On for 1 minute unless turned off manually)

Battery: 6 x AA (Battery life below based on alkaline type)

Battery life: Approx. 135 hours (without backlight and wireless)

Low battery indication:

is displayed when the battery

voltage drops below the operating level

Auto shut off: 30 minutes of inactivity when APO is activated

Over range: "OL" or "-OL" is displayed

Operating environment: 32°F to 122°F (0°C to 50°C) at <75% RH

Temperature coefficient: 0.1 x (specified accuracy) per °C (0°C to

18°C, 28°C to 50°C), per 0.6°F (32°F to 64°F, 82°F to 122°F

Storage temperature: -4°F to 140°F (-20°C to 60°C), 0 to 80% RH

(with battery removed)

Weight: 4.03 lbs (1.83 kg)

Pressure

Sensor type: Absolute pressure sensors

Connector type: Standard 1/4" and 3/8" NPT male flare fitting

Range: 29” HgV to 580psig (English), 74 cmHgV to 0 to 4000KPa

(Metric), 4.000MPa (Metric), and 40.00bar (Metric)

Resolution: 0.1 psi/inHg; 1 kPa/cmHg; 0.001MPa; 0.01bar

Accuracy: 29” HgV to 0” HgV: ±0.2” HgV

74 cmHgV to 0 cmHgV: ±1 cmHgV

0 to 200 Psig: ±1 Psi; 0 to 1378 KPa: ±7 KPa; 0 to 1.378MPa:

±0.007MPa; 0 to 13.78 bar ±0.07 bar

200 to 580 Psig: ±0.3%+1 Psi; 1378 to 4000KPa: ±0.3%+7 Kpa;

1.378 to 4.000MPa: ±0.3%+0.007MPa; 13.78 to 40.00bar:

±0.3%+0.07bar

Maximum overload pressure: 800 psig

Units: Psig, kPa, MPa, bar, inHg, and cmHg

Microns for Vacuum

Connector type: Standard 1/4" and 3/8" NPT male flare fitting

Range: 50 to 9999 microns of mercury

Resolution: 1 micron (50 to 2000 microns), 250 microns (2001 to

5000 microns), 500 microns (5001 to 8000 microns), 1000 microns

(8001 to 9999 microns)

Accuracy: ±10% or ±10 microns, whichever is greater (50 to 1000

microns)

Maximum overload pressure: 580 psig

Units: Microns of mercury

Temperature

Sensor type: Type K thermocouple

Range: -95°F to 999.9°F (-70°C to 537.0°C)

Resolution: 0.1°F/°C

Accuracy: ±(1.0°F) -95°F to 199.9°F; ±(0.5°C) -70°C to 93°C

±(2.0°F) 200°F to 999.9°F; ±(1.0°C) 93°C to 537.0°C

Note: All accuracies are after a field calibration.

Wireless

Wireless range: 1 to 100 feet (30m)

Initial wireless sync range: More than 1 foot and less than 10 feet

Wirelessly transmit to Fieldpiece HG3

Wirelessly receive indoor wet bulb (IDWB) and outdoor dry bulb

(ODDB) from Fieldpiece transmitters

Note: If the indoor wet bulb or outdoor dry bulb measurements result

in an uncalculable TSH, "OL" or "-OL" will display in the lower left
corner.

Note: If any one of the parameter IDWB or ODDB is synced wirelessly

to SMAN4 then the other parameter ODDB or IDWB has to be
synced wirelessly too.

Maintenance

Clean the exterior with a dry cloth. Do not

use liquid.

Battery Replacement

The battery must be replaced when the

battery life indicator

is empty. Turn your

SMAN4 off and replace with 6 AA batteries.

Cleaning the Micron Sensor

Over time, the vacuum sensor of the SMAN4

may become contaminated with dirt, oil, and

other contaminants introduced from pulling

vacuums.

1 Never use an object such as a cotton swab to clean the sensor,

you may cause damage to the sensor.

2 Open all knobs/valves, and cap all the ports except for the VAC

port.

3 Drop enough Isopropyl (rubbing) alcohol into the VAC port using

an eye dropper or funnel so that it can flush out contaminants.

4 Cap VAC port and gently shake your SMAN4 upside down to clean

sensor.

5 Turn right side up. Open a port to pour out the rubbing alcohol

and open all ports to allow sensors to dry out. Drying usually

takes about an hour.

Calibration

Temperature

To calibrate your SMAN4 temperature

thermocouples, adjust the pot on the front of

the meter labeled T1 Cal or T2 Cal. The best way

to calibrate is to match to a known temperature.

Ice water is very close to 32°F and is readily

available. Accuracies of one degree or better

are obtained through field calibration.

1 Stabilize a large cup of ice water by stirring. Pure, distilled water

will be the most accurate.

2 Press the circular button until your SMAN4 enters Direct

Temperature (T1 T2) mode.

3 Immerse the temp probe in ice water from T1 and adjust the

T1 Cal pot with a flat head screwdriver and let it stabilize, keep

stirring.

4 Repeat Step 3 for temp probe in T2.

Pressure Zeroing

To calibrate your SMAN4 pressure sensors to

atmospheric pressure, ensure that your SMAN4

is disconnected from any pressure source and at

equilibrium with the ambient pressure.

1 SMAN4 must be in Superheat/Subcooling, Saturation, or T1 T2

Direct mode.

2 Press the CAL Atmospheric Pressure button and your SMAN4 will

set the zero point of pressure to the ambient pressure.

Advanced Pressure Calibration

Your SMAN4 has the ability to perform a linear

adjustment of the pressure sensors based on

refrigerant type, temperature, and pressure.

Calibration setup: For best results, perform both the Temperature

and Pressure Zeroing procedures. See Calibration section for

details. This will ensure pressure readings are zeroed and

thermocouple is properly calibrated to the T1 port of the SMAN.

Calibration to T2 port is not necessary for this calibration.

The refrigerant cylinder should be stored in a stable ambient

environment for at least 24 hours before calibration.

1 Press circular button until your SMAN4 enters T1 T2 Direct mode.

2 Plug in a Type K thermocouple into T1. (A bead type

thermocouple, like the ATB1, is recommended.)

3 Connect the SMAN4 to a refrigerant cylinder of a known, single

refrigerant using an EPA approved service hose. Be sure to open

both HIGH and LOW side valves on your manifold and cap the

unused ports. (If caps are not available you can connect both

ends of a refrigerant hose to the two unused caps. Note: Some

refrigerant will remain in the hoses which will need to be

recovered.)

4 Press the REFRIGERANT button to match the refrigerant of the

cylinder you are using.

5 Attach bead-type thermocouple to the side of the cylinder using

tape. It is recommended to attach in the middle of the cylinder.

Important: Let the temperature of the thermocouple stabilize to

the refrigerant temperature for 1 to 2 minutes or until stable.

6 Open the refrigerant cylinder. The pressure inside cylinder should

now be displayed on both HIGH and LOW side pressure sensors.

7 Press the CAL Test Pressure button. If successful, "Good" will

display for 3 seconds. If failed, "Err" will display for same time.

Your SMAN checks with its built-in P-T charts to compare the

temperature of the refrigerant in the tank to the vapor saturation

temperature based on the refrigerant you selected. If the measured

pressures on your SMAN are within ±3psi of the P-T chart pressure

corresponding to the vapor saturation temperature, the SMAN will

adjust the pressure sensor linearity to match the P-T chart.

Possible causes of failed "Err" pressure calibration:

1. Refrigerant tank was not stored in stable ambient conditions for at

least 24 hours.

2. Thermocouple attached to refrigerant tank was not properly

calibrated to T1 port of SMAN.

3. Thermocouple was plugged into wrong port T2 instead of T1.

4. Incorrect refrigerant was selected on the SMAN.

Refrigerants

The P-T charts of the following 44 refrigerants come pre-programmed

into your SMAN4. In your SMAN4 the refrigerants are listed in order of most

commonly used. Here, they are listed in numerical order for easy reference.

R11, R113, R114, R12, R123, R1234YF, R124, R125, R13, R134A, R22,

R23, R401A(MP39), R401B, R402A, R402B, R404A, R406A, R407A,

R407C, R407F, R408A, R409A, R410A, R414B (Hotshot), R416A, R417A,

R417C (HOT SHOT 2), R420A, R421A, R421B, R422A, R422B(NU22B),

R422C(Oneshot), R422D, R424A, R427A, R434A(RS-45), R438A(MO99),
R500, R502, R503, R507A, R508B (Suva95)

Limited Warranty

This meter is warranted against defects in

material or workmanship for one year from

date of purchase. Fieldpiece will replace or

repair the defective unit, at its option, subject

to verification of the defect.

This warranty does not apply to defects

resulting from abuse, neglect, accident,

unauthorized repair, alteration, or unreasonable

use of the instrument.

Any implied warranties arising from the sale of

a Fieldpiece product, including but not limited

to implied warranties of merchantability and

fitness for a particular purpose, are limited to

the above. Fieldpiece shall not be liable for loss

of use of the instrument or other incidental or

consequential damages, expenses, or economic

loss, or for any claim of such damage, expenses,

or economic loss.

State laws vary. The above limitations or

exclusions may not apply to you.

Obtaining Service

Call Fieldpiece Instruments for current costs

on our fixed price warranty service. Send

check or money order for the amount quoted.

Send your digital manifold, freight prepaid, to

Fieldpiece Instruments. Send proof of date and

location of purchase for in-warranty service.

The meter will be repaired or replaced, at the

option of Fieldpiece, and returned via least cost

transportation.

For international customers, warranty for

products purchased outside of the U.S. should

be handled through local distributors.

www.fieldpiece.com

© Fieldpiece Instruments, Inc 2013; v25

!

WARNINGS

DO NOT APPLY MORE THAN 800 PSI TO ANY PORT ON THE MANIFOLD.

FOLLOW ALL EQUIPMENT MANUFACTURER'S TESTING PROCEDURES

ABOVE THOSE IN THIS MANUAL IN REGARDS TO PROPERLY

SERVICING THEIR EQUIPMENT.