Quick Start
Receive a Wireless Reading
1. Select RECV on main dial.
2. Hold SYNC for one second.
3. Set partner device to transmitting switch
position.
4. Hold SYNC on partner device for one second.
When connection is made
and
icons
will display on the top of LT17AW screen.
5. Enjoy the wireless goodness!
Certifications
IEC/EN61010-1
EMC EN61326-1
FCC
C-Tick (N22675)
WEEE
RoHS compliant
CATIII: Is for measurements performed in the
building installation.
Description
As a proud owner of the first wireless digital
multimeter, LT17AW, you are well on your way to
eliminating tangled wires. LT17AW will give you
the power to receive measurements wirelessly
from anywhere on the jobsite. For instance, you
can transmit indoor wet bulb measurements
over-the-air while you work at the condenser.
Your LT17AW can receive measurements
wirelessly from any Fieldpiece accessory head
connected to a Fieldpiece wireless transmitter
like ET2W or EH4W.
In addition, your LT17AW can also transmit
wirelessly any measurement range on the dial
to any Fieldpiece wireless receiver, like EH4W,
HG3, or another LT17AW.
Your LT17AW easily syncs with compatible
wireless Fieldpiece transmitters and receivers.
LT17AW has a wide wireless range to cover the
whole jobsite.
Your LT17AW remembers the last meter it was
synced with, so there is no need to constantly
resync the same pairs. Note: this feature is
available when synched to single-link receivers
such as the LT17AW, ET2W, and EH4W.
Controls
Rotate dial to the function you want to use.
Hold to connect to other Fieldpiece
wireless products.
Press to activate MAX or MIN. Hold to
return to normal mode.
Press to freeze reading on display.
Press to toggle backlight on and off.
Fieldpiece
Wireless
Digital Multimeter
OPERATOR'S
MANUAL
Model LT17AW
Display
Battery Life (replace 9V if blinking)
O
F
F
MFD
NCV
V
AC
V
DC
A
AC
Hz
Hz
%
°C
°F
SC56
AUTO OFF
TRUE RMS
CAT III
600V
TEMP
K-TYPE
30V
MAX
SEL
HOLD
INRUSH
Auto Power Off Enabled
Wireless Icon
Singal Strength Bars
k M
MAXMIN HOLD
LT17AW
EH4W
HOLD (freezes current reading)
k M
MAXMIN HOLD
LT17AW
EH4W
MAX (displays maximum reading)
k M
MAXMIN HOLD
LT17AW
EH4W
MIN (displays minimum reading)
O
F
F
MFD
NCV
V
AC
V
DC
A
AC
Hz
Hz
%
°C
°F
SC56
AUTO OFF
TRUE RMS
CAT III
600V
TEMP
K-TYPE
30V
MAX
SEL
HOLD
INRUSH
Continuity Test
O
F
F
MFD
NCV
V
AC
V
DC
A
AC
Hz
Hz
%
°C
°F
SC56
AUTO OFF
TRUE RMS
CAT III
600V
TEMP
K-TYPE
30V
MAX
SEL
HOLD
INRUSH
Kilo Unit (10
3
, one thousand)
O
F
F
MFD
NCV
V
AC
V
DC
A
AC
Hz
Hz
%
°C
°F
SC56
AUTO OFF
TRUE RMS
CAT III
600V
TEMP
K-TYPE
30V
MAX
SEL
HOLD
INRUSH
Mega Unit (10
6
, one million)
Specifications
Wireless range: Up to 75 feet (23m) line of sight
Minimum Wireless Distance: Up to 1 foot (30cm)
Wireless Frequency: 910MHz to 920MHz (US), 868.1MHz to
868.5MHz (European)
Display: 5000 count liquid crystal display with backlight
Overrange: (OL) or (-OL) is displayed Zero: Automatic
Measurement rate: 2.5 times per second, nominal
Operating environment: 32°F to 122°F (0°C to 50°C) at <70% RH
Storage temperature: -4°F to 140°F (-20°C to 60°C), 0 to 80% RH
(with battery removed)
Accuracy: Stated accuracy @ 73°F±9°F (23°C±5°C), <75%RH
Temperature coefficient: 0.1 x (specified accuracy) per °C [0°C to
19°C (32°F to 66°F), 28°C to 50° C(82°F to 122°F)]
APO (Auto Power Off): Approx. 30 minutes. Disable APO by
holding the MAX/MIN button while turning on the meter. APO
will not show on the LCD. APO is disabled in any wireless mode.
Backlight: Bright blue backlight. Approx 3 minutes.
Power: Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22
Battery life: 100 hours typical alkaline
Low battery indication: Battery icon will be empty (
). "bAtt"
displays along with a continuous beep when the battery voltage
drops below the operating level. Meter shuts off in 5 seconds.
Dimensions: 165mm(H) x 78mm(W) x 50mm(D)
Weight: Approx. 315g including rubber boot
Overload protection: 600 VDC or 600VAC rms unless otherwise
stated. 15 seconds on 500mV range.
Test leads: Use the test leads that are supplied with the meter or UL
listed test leads rated CATIII 600V or above.
Safety: Designed to meet IEC 61010-1(EN 61010-1), CAT III 600V,
Class II, polution degree 2, indoor use and comply with CE.
UL61010-1.
Functions
Wireless Receiver Mode
Use your LT17AW to wirelessly receive
measurements from any Fieldpiece wireless
transmitter you have set up on the jobsite.
LT17AW
1. Select RECV switch position. Hold SYNC but-
ton for one second. Search pattern initiates.
Fieldpiece Transmitters
2. Connect any Fieldpiece accessory head to a
Fieldpiece transmitter.
3. Select DC switch position for all accessory
heads except ACH4 (AC switch position).
4. Hold SYNC button for one second.
Note: When a connection is made, the signal
strength and battery life of the partner de-
vice will be displayed on top of the LT17AW
screen. If no partner device is found within
two minutes, the LT17AW will stop searching.
Wireless Transmitter Mode
Transmit measurements wirelessly direct
from LT17AW or from an attached Fieldpiece
accessory head to Fieldpiece wireless receivers,
which lets you be in multiple places at once.
LT17AW Sending Direct Measurements
1. Select any switch position other than RECV.
2. Hold SYNC for one second until meter beeps.
Search pattern will initiate on LCD.
Fieldpiece Receivers
3. Select RECV switch position. Hold SYNC but-
ton for one second. For HG3, highlight the test
line and hold SYNC button for one second.
Note: When a connection is made, the signal
strength to the partner device will display on
LT17AW. If no partner device is found within
two minutes, the LT17AW will stop searching.
LT17AW Transmitting Accessory Head Reading
1. Connect any Fieldpiece accessory head to the
LT17AW by removing the probe tips from test
leads and plugging into accessory head plugs.
Insert test leads to COM and VΩ ports.
2. Select a mVDC range for all accessories except
for ACH4 (mVAC range).
3. Hold the SYNC button for one second.
4. Follow "Fieldpiece Receivers" section above.
Wireless Auto-Connection
When powered on, LT17AW will search
and connect to the last connected single-link
partner device (LT17AW, ET2W, or EH4W).
Temperature (°F/°C)
Plug any K-type thermocouple directly into the meter to measure
temperature. Cold junction is located “inside meter” and allows for
extremely accurate measurements even in rapidly changing ambient
temperatures (going from rooftop to freezer). No adapter is required.
See Temp Calibration section for calibration instructions.
Range: -30°F to 1400°F, (-35°C to 750°C) Resolution: 0.1°
Accuracy: ±(1°F) 32°F to 120°F, ±(1°C) 0°C to 49°C
±(1%+1.5°F) 32°F to 750°F, ±(1%+1°C) 0°C to 399°C
±(3%+4°F) -30°F to 32°F, ±(3%+3°C) -35°C to 0°C
±(3%+4°F) 750°F to 1400°F, ±(3%+3°C) 400°C to 750°C
Sensor type: K-type thermocouple
Overload protection: 30 VDC or 30VAC rms
Voltage AC (VAC) (50Hz-500Hz)
Test power lines (120, 220, 480), test 24V going to controls, and
test for transformer failure.
Ranges: 500mV, 5000mV, 50V, 500V, 600V Resolution: 0.1mV
Accuracy: ±(1.2% + 6) 500mV to 50V ranges
±(2%+6) 500V and 600V ranges
Input Impedance: 10MΩ
Voltage DC (VDC)
Select VDC and the range will automatically be selected to give
the best resolution.
Ranges: 500mV, 5000mV, 50V, 500V, 600V Resolution: 0.1mV
Accuracy: ±(0.5% + 3)
Input impedance: 10MΩ
FUSED
200
m
A
MAX
FUSED
30V
MAX
WIRELESS
ENABLED
MAX/MIN
SYNC
HOLD
CAT.III
MAX
600V
15 SEC FOR
500
m
V
RANGE
AUTO-OFF
500
600
600
50
5000m
500m
500µ
50m
200m
200m
500m
5000m
50m
50500
500µ
500
750
500
1400
200
2
K
200
20
M
2
K
200
K
20
RECV
k M
MAXMIN HOLD
LT17AW
EH4W
500
600
600
50
5000m
500m
500µ
50m
200m
200m
500m
5000m
50m
50500
500µ
500
750
500
1400
200
2
K
200
20
M
2
K
200
K
20
RECV
FUSED
200
m
A
MAX
FUSED
30V
MAX
CAT.III
MAX
600V
15 SEC FOR
500
m
V
RANGE
Unplug Leads and Slide
TEMP Switch to the Left
500
600
600
50
5000m
500m
500µ
50m
200m
200m
500m
5000m
50m
50500
500µ
500
750
500
1400
200
2
K
200
20
M
2
K
200
K
20
RECV
500
600
600
50
5000m
500m
500µ
50m
200m
200m
500m
5000m
50m
50500
500µ
500
750
500
1400
200
2
K
200
20
M
2
K
200
K
20
RECV
500
600
600
50
5000m
500m
500µ
50m
200m
200m
500m
5000m
50m
50500
500µ
500
750
500
1400
200
2
K
200
20
M
2
K
200
K
20
RECV
MAX/MIN
SYNC
HOLD
MAX/MIN
SYNC
HOLD
MAX/MIN
SYNC
HOLD
MAX/MIN
SYNC
HOLD