Microamp and

Milliamp
Accessory Head
Model: AUA2

Description

The AUA2 microamp and milliamp accessory

head converts microamps (for flame rectifier diode)
and milliamps to a mV signal for direct current read-
ing on any meter with mV ranges. For resolution of
0.1 micro/milliamps, use a meter that displays
0.1mV.

The AUA2 along with the included quick connect

and mini-plug adapters converts the Fieldpiece
"stick" series meter, DL3 data logger, and EHDL1
electronic handle to a one-piece flame rectifier
diode tester. Use the optional Fieldpiece ADLS2
deluxe silicone test leads or the AHDL1 adapter
handle with the AUA2 for use with DMMs with
industry standard jacks.

EHDL1 AHDL1w/DMM DL3 “Stick”

Operation

1. Connect the AUA2 to meter’s COM and VOLTS

plugs. Connect through leads or slide head onto
meter like shown to the left.

2. Break flame rectifier diode circuit and connect

AUA2 in series using the appropriate adapter.
You can also use industry standard leads.

3. Select microamps (for flame rectifier diode) or

milliamps on the front of AUA2.

4. Set the meter to mV AC or DC

range to directly read

micro/milliamps.

Specifications

Conversion rate: 1mV/1mA, 1mV/1μA
Resolution: 0.1 for meters with 0.1mVDC
Accuracy: (at 73ºF ± 10ºF 95%RH)

DC: ±(1.0%rdg + 1dgt) AC: ±(1.5% + 2dgts)

Max current: 200mA AC or DC
Overload protection: 0.25A/600V (6x32mm) fast

blow fuse, Fieldpiece part # RFM66.

Operating temp: 32ºF to 120ºF at <95%RH
Storage temp: -4ºF to 140ºF
Safety: CE- mark approved, EN61010-1 approval,

Cat III 600V

Warranty

The product is warranted to the original purchas-

er against defects in material or workmanship for a
period of one (1) year from the date of purchase.
During the warranty period, Fieldpiece Instruments
will, at its option, replace or repair the defective unit.

This warranty does not apply to defects resulting

from abuse, neglect, accident, unauthorized repair,
alteration, or unreasonable use of the instrument.
Any implied warranty arising out of the sale of
Fieldpiece's products including but not limited to
implied warranties of merchantability, and fitness for
purpose, are limited to the above. Fieldpiece shall
not be liable for incidental or consequential dam-
ages.

Service

Return any defective AUA2 to Fieldpiece for war-

ranty service along with proof of purchase. Contact
Fieldpiece for out of warranty repair charges.

Fieldpiece Instruments, Inc.

California, U.S.A.

www.fieldpiece.com

Microamp and

Milliamp

Accessory Head
Model: AUA2

Description

The AUA2 microamp and milliamp accessory

head converts microamps (for flame rectifier diode)
and milliamps to a mV signal for direct current read-
ing on any meter with mV ranges. For resolution of
0.1 micro/milliamps, use a meter that displays
0.1mV.

The AUA2 along with the included quick connect

and mini-plug adapters converts the Fieldpiece
"stick" series meter, DL3 data logger, and EHDL1
electronic handle to a one-piece flame rectifier
diode tester. Use the optional Fieldpiece ADLS2
deluxe silicone test leads or the AHDL1 adapter
handle with the AUA2 for use with DMMs with
industry standard jacks.

EHDL1 AHDL1w/DMM DL3 “Stick”

Operation

1. Connect the AUA2 to meter’s COM and VOLTS

plugs. Connect through leads or slide head onto
meter like shown to the left.

2. Break flame rectifier diode circuit and connect

AUA2 in series using the appropriate adapter.
You can also use industry standard leads.

3. Select microamps (for flame rectifier diode) or

milliamps on the front of AUA2.

4. Set the meter to mV AC or DC

range to directly read

micro/milliamps.

Specifications

Conversion rate: 1mV/1mA, 1mV/1μA
Resolution: 0.1 for meters with 0.1mVDC
Accuracy: (at 73ºF ± 10ºF 95%RH)

DC: ±(1.0%rdg + 1dgt) AC: ±(1.5% + 2dgts)

Max current: 200mA AC or DC
Overload protection: 0.25A/600V (6x32mm) fast

blow fuse, Fieldpiece part # RFM66.

Operating temp: 32ºF to 120ºF at <95%RH
Storage temp: -4ºF to 140ºF
Safety: CE- mark approved, EN61010-1 approval,

Cat III 600V

Warranty

The product is warranted to the original purchas-

er against defects in material or workmanship for a
period of one (1) year from the date of purchase.
During the warranty period, Fieldpiece Instruments
will, at its option, replace or repair the defective unit.

This warranty does not apply to defects resulting

from abuse, neglect, accident, unauthorized repair,
alteration, or unreasonable use of the instrument.
Any implied warranty arising out of the sale of
Fieldpiece's products including but not limited to
implied warranties of merchantability, and fitness for
purpose, are limited to the above. Fieldpiece shall
not be liable for incidental or consequential dam-
ages.

Service

Return any defective AUA2 to Fieldpiece for war-

ranty service along with proof of purchase. Contact
Fieldpiece for out of warranty repair charges.