5 Troubleshooting 

5.1 General

For a correct analysis of the proper operation of the MASS‐VIEW

®

  meter it is recommended to remove 

the unit from the process line and check it without applying gas supply pressure. In case the unit is dirty, 
this  can  be  ascertained  immediately  by  loosening  the  couplings.  Furthermore  check  if  the  RJ‐45 
connector is fixed properly. Energizing or de‐energizing of the instrument indicates whether there is an 
electronic failure. The OLED screen should light up. 
After that, fluid pressure is to be applied in order to check behaviour. 
If there is suspicion of leakage, do not check for bubbles with a leak detection liquid under the cover as 
this may lead to a short‐circuit at the sensor or PC‐board. 

5.2 Troubleshooting

summary

Symptom 

 

Possible cause

 

Action

 

No output signal, no display, no flow indication 

No power supply 

1a) Check power supply 

 

 

1b) Check cable(s) 

 

PCB failure 

1c) contact online: Mass Flow ONLINE* 

 

Valve blocked 

1d) contact online: Mass Flow ONLINE* 

 

Filter blocked 

1e) Clean filter  

 

Sensor failure 

1f) contact online: Mass Flow ONLINE* 

Maximum output signal, maximum flow 

indication 

Flow too high 

2a) reduce the flow or re‐range the 

instrument 

 

PCB failure 

2b) contact online: Mass Flow ONLINE* 

 

Sensor failure 

2c) contact online: Mass Flow ONLINE* 

High setpoint vs. output offset, flow indication 

lower than expected 

Pressure not correct 

3a) Check pressure 

 

Blockage in line 

3b) Remove cause  

 

Valve blocked 

3c) contact online: Mass Flow ONLINE* 

 

PCB failure 

3d) contact online: Mass Flow ONLINE* 

Signal or flow indication lower than expected 

System stoppage 

4a) Remove cause  

Flow is gradually decreasing or changing 

(Gas) Condensation 

5a) Decrease supply pressure or increase 

temperature 

 

Pressure  has changed 

5b) Check valve setting 

Small flow occurs when valve is supposed to be 

closed 

Valve is leaking 

6) contact online: Mass Flow ONLINE* 

 
 

 

 

 

 

 

 

 

For troubleshooting issues or returns, please go to 

www.massflow‐online.com

 and click on service. 

Here you find the instructions how to fix the problem (FAQ’s) or how to return an instrument (RMA). 

 

5.3 Service

and

repair

 

 

MASS‐VIEW

®

  instruments can be repaired but not all the parts. Parts that can be repaired can be 

found on the website of Mass Flow ONLINE (

www.massflow‐online.com