3 Operation 

3.1 General

Bronkhorst  High‐Tech  instruments  are  designed  in  such  a  way  that  they  will  meet  user  process 
requirements  in  the  best  possible  way.  Flow  meters  and  controllers  are  powered  from  a  dc  power 
source with a minimal ripple. When providing your own power supply be sure that voltage and current 
rating are according to the specifications of the instrument and furthermore that the source is capable 
of delivering enough energy  to the instrument. Cable wire diameters should  be sufficient to carry the 
supply current and voltage losses must be kept as low as possible. 

3.2 Power

and

warm-up

Before  switching  on  the  power,  be  sure  all  connections  have  been  made  according  to  the  hook‐up 
diagram in 2.7. Check gas connections and make sure there is no leakage. If needed purge the system 
with a proper gas. Turn on power and allow at least 30 minutes for warming up and stabilizing. During 
the warm‐up period, gas pressure may either be on or off.  

 

3.3 Setting

up

the

instrument

After powering on, the first time, the instrument will ask which gas is used and which flow range is to be 
used (see picture below).  
Navigate to the fluidname using the 4‐way button (up and down direction) and press in the centre to 
make a selection, then select the range using the up and down direction and press also in the centre to 
make the selection. 
 
After selecting the fluid and the range, the communication type can be selected .   
The settings will be stored when the “save” button is pressed. 

 
Then  turn  on  gas  supply  gently.  Avoid  pressure  shocks,  and  bring  the  instrument  gradually  up  to  the 
level of the actual operating conditions. Also switch off gas supply gently. 
 
The  MASS‐VIEW

®

    is  designed  in  such  a  way  that  for  a  change  in  the  range  or  gas,  you  simply  select 

another available range or gas in the menu. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

After entering the default password (which is ‘abc’) in the security menu, all menu’s can be enabled or 
disabled. Also a new password can be entered and saved.