Bronkhorst MASS-VIEW Series User Manual

Page 11

background image

 

BRONKHORST HIGH‐TECH 

 

 

 

9.17.051 

MASS‐VIEW® series 

Page 11 

2.3 Mounting

The mounting position of the MASS‐VIEW

®

  is vertical, although it is possible to turn the device 180° and 

change  the  display  setting.  Avoid  installation  in  close  proximity  of  mechanic  vibration  and/or  heat 
sources. 

 

 

2.4 In-line

filter

2.5 Fluidic

connections

The  MASS‐VIEW

®

    can  be  equipped  with  (compression)  fittings.  Tighten  the  nut  finger  tight;  while 

holding  the  instrument,  then  tighten  the  nut  1  turn.  If  applicable  follow  the  guidelines  of  the 
manufacturer  of  the  fittings.  While  tightening  fittings,  do  not  apply  excessive  force,  in  order  to  avoid 
damaging in/output thread or other sensitive parts of your instruments. 
  
Please be aware when using Oxygen (O

2

) that the fittings mounted to the body should be absolutely 

oxygen clean. 

2.6 Piping

Be sure that piping is absolutely clean! 
 

 

DO  NOT  install  small  diameter  piping  on  high  flow  rates,  because  the  inlet  jet  flow  will  affect  the 
accuracy.  
DO NOT mount abrupt angles direct on in‐ and outlet, especially not on high flow rates. At least 10 pipe 
diameters distance between the angle and the instrument is recommended. 
DO NOT mount pressure regulators direct on the inlet of gas flow meters/controllers, but allow some 
meters of piping (at least 25 pipe diameters).  

2.7 Electrical

connections

The MASS‐VIEW

®

  instruments are powered with +15 Vdc to +24 Vdc. 

When  providing  your  own  power  supply  be  sure  that  voltage  and  current  rating  are  according  to  the 
specifications  of  the  instrument(s)  and  furthermore  that  the  source  is  capable  of  delivering  enough 
power to the instrument. Refer to Hook‐up MASS‐VIEW

®

  series   (document  nr.  9.16.085)  for  more 

details. 

 

When  using  other  cables,  cable  wire  diameters  should  be  sufficient  to  carry  the  supply  current  and 
voltage losses must be kept as low as possible.  

 

 

When connecting the system to other devices (e.g. to PLC), be sure that the integrity of the shielding is 
not affected. Do not use unshielded wire terminals. 

 

Although the gases to be measured should be absolutely free of dirt, oil, moisture and other particles, 
it is recommended to install an in‐line filter upstream of the flow meter / controller, and if backflow 
can  occur,  a  downstream  filter  is  recommended  too.  Be  aware  of  the  pressure  drop  caused  by  the 
filter. 

 

Always  check  your  system  for  leaks,  before  applying  fluid  pressure.  Especially  if  toxic,  explosive  or
other dangerous fluids are used. 

 

We recommend using our standard  cables. These cables have the right connectors and if loose ends 
are used, these are coloured and marked to prevent improper connection. 
 
All cabling options can be purchased online 

www.massflow‐online.com