3.4.4 Sleep

mode

To increase the lifetime of the display a sleep mode is built‐in which displays the flow at a random place 
on the screen. The sleep mode will become active when the readout screen is visible and no button is 
pressed for a number of minutes.  
The screensaver and the delay can be enabled in the settings menu. 
 
In addition to the sleep mode the display contrast can be adjusted in 5 steps in the settings menu. 

3.4.5 Counter 

The counter in the MASS‐VIEW

®

  can be set up in the “counter” menu and has 3 different modes which 

can be selected via the “mode” parameter in the “counter” menu: 
1. Off 

: counter disabled. 

2. Up  

: counting upwards continuously.  

3. Up to..  : counting up to limit (batch counter). 
 
When the counter mode is set to “Up” or “Up to..” the following parameters will be displayed 
 
Unit:  

 The output unit in which the counter will be displayed on the screen. (FlowDDE parameter 
123 or 128). 

Limit:  

Batch limit. This parameter is only active when counter mode “Up to..” is selected. 

Reset:  

Resets the counter. 

3.4.6 Alarms 

The alarms can be adjusted in the Alarms menu using the following parameters: 
 
Activate (FlowDDE parameter 118) 
This parameter sets one of the following alarm modes. 
 

Name: 

DDE Value:

Description:

None 

Alarms Off.

Min + max 

Alarm on absolute limits.

Response 

Alarm on limits related to setpoint/marker. 

 
When the alarm mode is set to “None”, the other alarm parameters are not visible in the menu. 
 
Set max / Set min (FlowDDE parameter 116 and 117) 
The minimum and maximum alarm values can be set between 0% and 130%. 
An alarm can be disabled by setting the value to 0%. 
When  alarm  mode  “min  +  max”  is  selected,  these  are  absolute  values,  when  “Response”  is  selected 
these values are relative to the setpoint or marker value. This enables the user to create a band around 
the setpoint/marker.  
 
Delay (FlowDDE parameter 182) 
The delay parameter is the time in seconds the alarm action will be delayed when alarm limit has been 
exceeded. This is also the time automatic reset will be delayed when sensor signal reaches the desired 
level again.