Configuration

2.1.3 Environmental Requirements

• For indoor use only
• Installation Category II (over voltage)
• Pollution Degree 2
• Altitude up to 2000 meters (with power derating)
• Relative Humidity 80% RH Max
• Ambient Operating Temperature 0-40 C

2.1.4 Rack

Mount

The SLM-4 chassis is designed to permit mounting in a standard 19-inch rack for system
application.

2.1.5 Tilt

Adjustment

The SLM-4 chassis is equipped with tilt stands on its two front feet, which can be flipped forward
to provide an inclined viewing angle of the front panel for bench-top use.

2.1.6

Service or Repair

If the instrument is damaged, please attach a tag to the instrument, identifying the owner and
indicating the required service. Follow the procedures detailed on Page iii of this manual.

2.1.7 Accessories

For each module configured, the following parts should be included in the shipment:

• Three-conductor 18 gauge power cord, 1 ea.
• SLM-4 electronic load chassis operation manual, 1 ea.
• Vsense cable(s) for each input
• Hook terminals for each input
• Banana plugs for each input
• Imonitor BNC cable (if applicable)

2.2 CONNECTIONS

2.2.1 GPIB

Connection

The GPIB connector on the rear panel (Figure 2–3) connects the to a controller and to other
GPIB devices. A GPIB system can be connected in any configuration (star, linear, or both), with
the following provisions:

• The maximum number of devices including the controller is no more than 15.
• The maximum length of all cables is no more than 2 meters times the number of

devices connected together, up to 20 meters maximum.

Ensure that the lock screws are firmly hand-tightened; use a screwdriver only for the removal of
screws. Figure 2–3 shows the rear panel of the SLM-4 chassis, where the GPIB connector is
located. The GPIB address of the SLM-4 chassis is set on the front panel (Figure 3–1).