Specifications, Dc voltage measurement, Dc voltage source – Fluke 725 User Manual
Page 66: Millivolt measurement and source

725
Users Manual
56
Specifications
Specifications are based on a one year calibration cycle
and apply from +18
°
C to +28
°
C unless stated otherwise.
All specifications assume a 5 minute warmup period.
DC Voltage Measurement
Range
Resolution
Accuracy,
(% of Reading + Counts)
30 V
(upper
display)
0.001 V
0.02 % + 2
20 V
(lower
display)
0.001 V
0.02 % + 2
90 mV
0.01 mV
0.02 % + 2
Temperature coefficient -10
°
C to 18
°
C, +28
°
C to
55
°
C:
±
0.005 % of range per
°
C
DC Voltage Source
Range
Resolution
Accuracy,
(% of Reading + Counts)
100 mV
0.01 mV
0.02 % + 2
10 V
0.001 V
0.02 % + 2
Temperature coefficient -10
°
C to 18
°
C, +28
°
C to
55
°
C:
±
0.005 % of range per
°
C
Maximum load: 1 mA
Millivolt Measurement and Source*
Range
Resolution
Accuracy
-10 mV to 75 mV
0.01 mV
±
(0.025 % + 1 count)
Maximum input voltage: 30 V
Temperature coefficient -10
°
C to 18
°
C, +28
°
C to
55
°
C:
±
0.005 % of range per
°
C
*Select this function by pressing
T
. The signal is
available at the thermocouple miniplug connector.