Auto Ramping the Output
Auto ramping gives you the ability to continuously apply a
varying stimulus from the calibrator to a transmitter, while
your hands remain free to test the response of the
transmitter.
When you press
L
, the calibrator produces a
continuously repeating 0 % - 100 % - 0 % ramp in your
choice of three ramp waveforms:
•
E
0 % - 100 % - 0 % 40-second smooth ramp
•
P
0 % - 100 % - 0 % 15-second smooth ramp
•
N
0 % - 100 % - 0 % Stair-step ramp in 25 %
steps, pausing 5 seconds at each step. Steps are
listed in Table 7.
To exit ramping, press any button.
Table 7. mA Step Values
Step
4 to 20 mA
0 %
4.000
25 %
8.000
50 %
12.000
75 %
16.000
100 %
20.000
Storing and Recalling Setups
You can store up to eight of your settings in a nonvolatile
memory and recall the settings for later use. A low battery
condition or a battery change does not jeopardize the
stored settings. Proceed as follows:
1. After you create a calibrator setup, press
S
. In the
display, the memory locations appear.
2. Press
Y
or
Z
to select locations one through eight. An
underscore appears below the selected memory
location.
3. Press
S
, only the stored memory location will be
displayed.
The setup is stored.
To recall setups, proceed as follows.
1. Press
Q
. The memory locations appear on the
display.
2. Press
Y
or
Z
to select the appropriate location and
press
Q
.